Description
The APNA-370 uses a combination of the dual cross flow modulation type chemiluminescence principle and the referential calculation method.
This gives it the advantages of the single-detector method plus the ability to do continuous measurements of NOX, NO, and NO2. The design gives great stability and extremely high sensitivity (0.1 ppm F.S.)
Standard equipment includes a drier unit with an automatic recycle function to provide dry ambient air as the ozone source. This makes long-term continuous measurements possible.
The detector uses a silicon photodiode sensor to reduce size and prolong working life.
All the necessary features are built right into a single rack-sized unit, including a reference-gas generator, an ozone-source drier unit, an ozone decomposer, and a sampling pump. No supplemental gas is required.
Specifications
Principle |
Cross flow modulation type, reduced pressure chemiluminescence (CLD) |
Application |
NO2, NO and NOX in ambient air |
Range |
Standard ranges: 0-0.1/0.2/0.5/1.0 ppm; auto range ~ manual range selectable; |
Repeatability |
±1.0% of F.S. |
Linearity |
±1.0% of F.S. |
Sample gas flow rate |
Approx. 0.8L/min |
Indication |
Measured value, range, alarm, maintenance screen |
Alarms |
During AIC, zero calibration error, span calibration error, temperature error in converter, etc. |
On-screen messages are available in four languages |
English, German, French, and Japanese. |
Input/output |
• 0-1 V/0-10 V/4-20 mA, to be specified (2 systems: either (1) momentary value and integrated or (2) moving average value) |
Power |
100/110/115/120/220/230/240 VAC, 50/60 Hz (to be specified) |
Dimensions |
430(W) x 550(D) x 221(H) mm |
Mass |
Approx. 20 kg, |