FEATURES
- Long-term stability : Cross-flow modulated non-dispersive infrared (NDIR) detection is renowned for long-term stability
- No need for optical adjustments
With NDIR, the sample gas and reference gas are intromitted into a single measurement cell alternately to obtain modulation signal. Therefore there is no need to adjust two different optical paths so that they are balanced. - A stable zero point
Since ENDA-5000 series output the difference between the measured gas and the reference gas each time measurement occures (once a second), the zero point is extremely stable. - Continuous cleaning keeps the cell clean
Since cleaning air is fed into the sample cell in between each batch of sample gas, the cell resistes contamination and normally remains clean.This reduces span drift and makes the equipment safe and stable for long periods of time.
- No need for optical adjustments
- Compact : Body yields wider maintenance area
ENDA-5000 series can be installed almost anywhere, with ample room on all sides for easy access and much easirer maintenance. - Easy to use : Features an intuitive touch panel
The operator can easily switch between the corrected and converted density settings screens or view alert information with touch of a single button. - Complies to world environemental regulations by local engineering
* Please consult HORIBA for further details.<Other Features>
- Continuous simultaneous measurement of up to 5 components with one system
HORIBA’s innovative optical technology enables ENDA-5000 series measure up to five components which can be arranged any combination. - Correction for interference
The interference correcting sensor uses a unique interference filter to compensate for the influence of interference by other gases. - Better alerts and extra alerts
In addition to the alert functions available in the past, the ENDA-5000 series feature extra alert functions. A continuous checking process can prevent the unit from stopping due to a failure, reducing the risk of failed measurements and assuring consistent operation. - Dramatically reduced correction time for SO2
Corrections of SO2 measurements using wet base methods of the past took a great deal of time (about 15 minutes), but with the ENDA-5000 series’ dry base method, correction takes only three minutes.
SPECIFICATIONS
Model | ENDA-5000 | ||||||
Component | NOx | SO2 | CO | CO2 | O2*1 | ||
Measurement methods | NDIR | NDIR | NDIR | NDIR | Magneto-pneumatic detection | ||
Range*2 | Standard | 200~5000 ppm | 200~5000 ppm | 200~5000 ppm | 5~25 vol% | 10~25 vol% | |
Option | 100 ppm~ | 50 ppm~ | 100 ppm~ | – | – | ||
Range Ratio | Within a factor of 10 | Within a factor of 10 | Within a factor of 10 | Within a factor of 5 | Within a factor of 2.5 | ||
Repeatability | Within 0.5% of full scale/week (with optional range, or during O2 measurement, ±1.0% of full scale/week) | ||||||
Linearity (indicator error) | ±1.0% of full scale | ||||||
Zero drift | ±1.0% of full scale/week (assuming surrounding temperature is maintained within 5 °C) (with optional range, or O2 measurement, ±2.0% of full scale/week) |
||||||
Span drift | ±2.0% of full scale/week (assuming surrounding temperature is maintained within 5°) | ||||||
Response time*3 | Within 60 seconds (Td + T90 from equipment intake area) (sample flow 0.6 L/min.) (within 240 seconds for SO2 only) | ||||||
Overall interference from co-existing gases | ±2.0% of full scale (within standard range, with standard gas formation) | ||||||
Display | Touch panel LCD (backlight) (four usable lines) | ||||||
Environment Condition |
Temperature | -5 to 40 °C (away from direct sunlight and radiation heat )*4 | |||||
Humidity | 85% or less (no condensation) | ||||||
Vibration | 100 Hz, 0.3 m/s2or less | ||||||
Dust | Standard environment or better | ||||||
Measuring Gas Condition | Temperature | 250 °C or lower | |||||
Dust | 0.1 g/Nm3 or less | ||||||
Standard gas composition*5 | NO: 500 ppm or less; NO2: 15 ppm or less; SO2: 1000 ppm or less; SO3: 50 ppm or less; CO: 200 ppm or less; CO2: 15% volume or less; H2O: 20% volume or less |
||||||
Sampling method | Dry sampling using an electric cooler | ||||||
Sample gas flow | 2.5 L/min~3.0 L/min | ||||||
Sample inlet tube | Teflon tubing (Ø8/Ø6 mm) | ||||||
Sample gas pressure | ±4.9 kPa (three points selected) (with no sample gas back pressure) |
|
|||||
Pressure control | Pressure control uses a regulator and pump; Reduced pressure sampling; Control pressure: -4.9 kPa | ||||||
Output | DC 4 to 20 mA (absolute output) (DC 0 to 16 mA/DC 0 to 1 V/DC 1 to 5V optional) Max. 12 output systems | ||||||
External output | Analysis alerts, analysis warnings, range display, corrections, conservation, purging (option) Contact capacity: DC 30 V 1 A, AC 250 V 1 A resistance load |
||||||
Correction method | Dry correction, automatic correction (correction cycle: 7 days standard, can be adjusted to between 1 and 99 days), manual correction | ||||||
Calibration gas | Zero gas | With measurement method authorization: N2, When there is no measurement method authorization: N2 or ambient air | |||||
O2 carrier gas | Ambient air | ||||||
Span gas | Gas cylinder for each component measured (when there is no measurement method authorization: O2 or ambient air can be used) | ||||||
Probe collection point filter | Flange: JIS 10K, 40 AFF; Sample probe tube length: 1000 mm; Material: SUS-316 stainless steel; Filter element: SUS-304 stainless steel and 2um-pleated quartz wool; Electric heater: 100 VA, with water droplet proof case |
||||||
Power supply | AC 100 V ±15 V(85 V~115 V) | ||||||
Power frequency | 50/60 Hz (switchable) | ||||||
Power consumption | About 800 VA (heated piping 30m: +1100 VA; heater in tray: +300 VA) | ||||||
Exterior dimensions/Mass | 600 (W) x 1770 (H) x 300 (D) mm (high pressure gas cylinders, 3.4 L cylinders, maximum of 3 cylinders); About 180 kg (not including cylinders) 600 (W) x 1770 (H) x 300 (D) mm (high pressure gas cylinders, 3.4 L cylinders, maximum of 3 cylinders); About 200 kg (not including cylinders) |
||||||
Materials in contract with sample gas | SUS-316 stainless steel, SUS-304 stainless steel, Teflon, polypropylene, polyethylene, flourorubber, PVC, PVDF, and glass |
||||||
Enclosure | Independent outdoor installation Plate thickness: Main unit, doors, top plate, steel plate: 2.3 mm; Channel base: 3.2 mm; Doors: front opening; Interface: right front |
||||||
Color/Finish | Semi-gloss Munsell 5Y7/1 on all inner and outer surfaces |