HORIBA HD-200 Industrial dissolved oxygen analyzer (Four-Wire Analyzer)

Realizes both the cartridge method membrane replacement and the membrane internal liquid replacement regeneration system Responds to a wide range of measurements (boiler water/rivers/aeration treatment, etc.)

In Stock
100% Original and quality
Extended Warranty

Description

FEATURES

  • Self-diagnosis of membrane damage and infiltration into the probe
  • On-site membrane replacement as well as whole sensor replacement
  • Span calibration by air preceded by auto-zeroing (Electrical auto zeroing function)
  • Extensive interfaces (2-channel transmission output, RS-485 compatible, 4-relay output)

SPECIFICATIONS

Measuring method Membrane type polarography method
Measuring range Dissolved oxygen concentration: 0 to 20 mg/L   Resolution: 0.01 mg/L
Saturation ratio: 0 to 200%   Resolution: 1%
Temperature: 0 to 50˚C   Resolution: 0.1˚C
Repeatability Dissolved oxygen concentration: ±0.5% full-scale or less, Temperature: ±0.5˚C (for equivalent input)
Linearity Dissolved oxygen concentration: ±0.5% full-scale or less, Temperature: ±0.5˚C (for equivalent input)
Transmission output Two points 4 to 20 mA DC   Input/output insulated type   Maximum load resistance 900 Ω
Contact output Three points Non-voltage contact output Relay contact, SPDT(1c)
Contact function R1, R2: Selectable from upper limit alarm, lower limit alarm, ON/OFF control, transmission output hold operation, cleaning output
FAIL: Malfunction alarm
Control operation Control range: 0.02 to 1.00 mg/L (±0.01 to ±0.50 mg/L)
Contact input One point   Contact format: Open collector non-voltage a contact
Contact function: Cleaning operation external input
Communication function RS-485   Two wire system, Input/output insulated type (not insulated for transmission output)
Temperature compensation range 0 to 50˚C
Ambient temperature -20 to 55˚C
Temperature measurement element Pt 1000 (0˚C) (Built into DO-1100 probe)
Calibration method Atmospheric calibration or saturated liquid calibration
Self-diagnosis function Calibration error, Sensor diagnosis error, Analyzer malfunction
Power supply 90 to 264 V AC 50/60 Hz
Power consumption 15 W or less
Construction Outdoor installation type: IP65 protection level
Mounting method: 50A pole or wall mounted
Case: Aluminum alloy
Mount fitting/hood: SUS304
Compatible sensor Sensor: 5505, 5510 Probe: DO-1100
Weight Approx. 4.5 kg
Equipment protected CE marking, FCC rules

TECHNOMASTER IS  A CERTIFIED DISTRIBUTOR FOR HORIBA IN EGYPT

Reviews

There are no reviews yet.

Be the first to review “HORIBA HD-200 Industrial dissolved oxygen analyzer (Four-Wire Analyzer)”

Your email address will not be published. Required fields are marked *